where:
l(mm), is the total length of the connecting traces expressed in millimeters.
This two-step tuned decoupling technique has been applied at 40.68 MHz for the coupling capacitors CC1, CC2 in Figs. 3 and 4, and the bypass capacitor CBP in Fig. 4. Knowing that the length of the viahole to the ground plane must be added to the connection length of a bypass capacitor (+1.6 mm for a standard FR4 epoxy glass board), two capacitors were selected with values of 5.6 nF and 3.3 nF. The table shows the electrical characteristics of these capacitors, and the associated values of the total length of the connecting traces calculated with Eq. 19.
From the table, one can see that with a required connection length of about 2 mm, a 5.6-nF chip capacitor is an optimum choice for the coupling capacitors CC1 and CC2, while a 3.3-nF chip capacitor that calls for a connection length of about 4 mm is well suited for performing a bypass to ground (capacitor CBP). It is important to note that the total length of the connecting traces calculated with Eq. 19 are approximate values, consequently the bonding pads of the capacitors have been enlarged of ±2 mm for performing an experimental tuning.
The second source of error mentioned above is calibration of the VNA. According to Agilent Technologies (Santa Rosa, CA),10 the best calibration technique at RF is the standard short-open-load-through (SOLT) calibration. If the test fixture is equipped with 3.5-mm SMA connectors, a model 85033D 3.5-mm calibration kit from Agilent Technologies is typically used. But this calibration kit moves the reference plane to the connector's outer conductor mating surface (plane P1 in Figs. 3 and 4), and not to the measurement plane (plane P2 in Figs. 3 and 4). Consequently, a second operation is needed to shift the calibration plane to the measurement plane. This operation can be performed with the VNA's port-extension function (mathematically modelling the test fixture) or with de-embedding software.10 Nevertheless, each method has limitations that produce additional errors. In order to avoid these errors, an efficient solution at RF is to move the calibration plane directly to the measurement plane using a technique called the direct measurement method.10
Direct measurement involves measuring physical calibration standards exactly in the same plane as the device under test (DUT) measurement plane and calculating error terms. With this method, the precise characteristics of the fixture do not need to be known. They are measured during the calibration process. From the start, the direct measurement method is a four-step process:
1. A user calibration (cal) kit must be built. Figure 6 shows the SOLT user cal kit used for the Philips' receiver. The short, open, and 50-Ω load are located at the end of a microstrip line having a length of 14 mm and 50-Ω characteristic impedance. Consequently, the through standard is a 2 × 14-mm long microstrip line having the same characteristic impedance.
2. The standards must be characterized: Fortunately, at RF, only the open standard requires characterization (because an unterminated microstrip line radiates energy). This characterization consists in determining the fringing capacitance of the open in the frequency band of interest. The method for measuring an open standard is detailed in ref. 10.
3. The standard definition table of the cal kit must be entered into the VNA's memory. With an Agilent VNA, this step is quickly performed by modifying an existing calibration kit.11
To minimize the number of standards to modify, it is wise to select a cal kit whose characteristics are as close as possible to those of the user kit. In the case of the kit shown in Fig. 6, the best solution is to modify a calibration kit in which the physical calibration standards are measured in the plane of calibration. The APC7 calibration kit matches this definition. With such a kit, the modification procedure is reduced to entering the fringing capacitance of the open standard:
When using an Agilent VNA, the fringing capacitance (CF) is modeled as a function of frequency by a cubic polynomial that best fits the fringing capacitance measured in step 2: